Product category:
Liquid Analysis: Titration, HPLC, IC
News Release from: Metrohm UK | Subject: Surfactrodes
Edited by the Processingtalk Editorial
Team on 26 December 2005
DIN/EN standard for determining anionic
surfactant
The determination of the content of surface-active agents (surfactants) plays an important role in many sectors - from wastewater analysis up to quality control in production processes
A new DIN/EN standard method has been announced for determining anionic surfactants by means of potentiometric two-phase titration The determination of the content of surface-active agents (surfactants) plays an important role in many sectors - from wastewater analysis up to quality control in production processes
This article was originally published on Processingtalk on 2 Jul 2008 at 8.00am (UK)
Related stories
Tighter process control in acid pickling baths
Online ion chromatography from Metrohm has enabled tighter process control of acid pickling baths, as the final cleaning stage prior to metal plating
The determination of acidity in soft drinks
Sparkling drinks are the main cause of increasing tooth erosion among teenagers: a study in the British Dental Journal found a strong link between fizzy drink consumption and tooth erosion
A method that is still in widespread use is the so-called Epton titration according to ISO 2271, a complicated and time-consuming manual titration.
The new DIN EN 14480 standard now describes an alternative determination method that is much simpler and faster: potentiometric two-phase titration.
Potentiometric two-phase titration according to DIN EN 14480: more precise, faster and much healthier! The solution of the anionic surfactant is pipetted into a titration vessel, to which a two-phase mixture of water and methyl isobutyl ketone/ethanol 1 : 1 and an emulsifier are added.
Under vigorous stirring the emulsion formed is titrated with a cationic surfactant, in this case benzethonium chloride (Hyamine 1622).
The ion pair formed, which is insoluble in water, is extracted immediately into the organic phase.
The determination of the equivalence point does not take place visually, but potentiometrically using surfactant-sensitive electrodes.
The Surfactrodes from Metrohm are used for this purpose; they have a high affinity for surface-active agents.
In comparison with the Epton titration you profit from the following advantages.
* Use of unproblematic solvents instead of chloroform.
* Short determination times: only a few minutes per titration.
* Objective, computer-supported determination of the equivalence point and therefore improved precision.
* Can easily be automated.
In an extensive Europe-wide interlaboratory test the potentiometric two-phase titration was compared with the Epton method and was found to yield the same analytical results with a better repeatability.
A detailed evaluation of the interlaboratory test can be found in the appendix of DIN EN 14480. Request a free brochure from Metrohm UK ...
• Metrohm UK: contact details and other news
• Email this article to a colleague
• Register for the free Processingtalk email newsletter
• Processingtalk Home Page


