News Release from: Metrohm UK | Subject: Voltammetry in Power plants
Edited by the Processingtalk Editorial
Team on 08 March 2007
Analysis of iron on power plants using
Voltammetry
In power plants Voltammetry can be used to determine the iron content in boiler waters, where Magnetite (an iron oxide) is used to help prevent corrosion: VA is thus an important corrosion indicator
Thermal power plants are at the core of the industrial society that we in the United Kingdom live in today and no matter whether the plant is nuclear of fossil fuel driven, its safe and reliable operation is critical and must be guaranteed Water is at the source of all thermal power plants as it enables cooling, expanding water vapour drives the turbines and water moderates the nuclear fission
This article was originally published on Processingtalk on 2 Jul 2008 at 8.00am (UK)
Related stories
Tighter process control in acid pickling baths
Online ion chromatography from Metrohm has enabled tighter process control of acid pickling baths, as the final cleaning stage prior to metal plating
The determination of acidity in soft drinks
Sparkling drinks are the main cause of increasing tooth erosion among teenagers: a study in the British Dental Journal found a strong link between fizzy drink consumption and tooth erosion
In fossil fuel and nuclear power plants a wide range of chemical reactions take place, some are related to normal operation but others provide important information about current or potential future faults.
Any metal corrosion of the metal components inside a power plant is caused by a reaction (pH dependant) between hydronium ions and oxygen reacting with the metal ions present.
Magnetite (a form of iron oxide) is used to keep the pH between 9.8-10.5 which keeps the surface of the boiler tubes active and helps prevent corrosion.
Further reading
The new Metrohm 856 conductivity module
The 856 Conductivity Module introduces the new 5-ring conductivity sensors which are far superior to the classical conductivity sensor
pH measurement of ethanol using the Aquatrode Plus
Investigations into the use of the Metrohm Aquatrode Plus electrodes for ethanol pH measurements according to ASTM D 6423 - the pH value is an important quality criterion for bioethanol
Determining organic additives in electroplating
ProcessLab-VA/CVS - the innovative process monitoring solution Voltammetry (VA) and Cyclic Voltammetric Stripping Analysis (CVS) are now available for routine analysis in the plant
Voltammetry (VA) can be used to determine the iron content in power plant waters, quantification of iron is important as this is a key corrosion indicator and allows any onsite maintenance to be pre-emptive rather than reactive which to the power plant operator is a much more costly scenario.
If there are appreciable iron deposits these can also inhibit heat transfer making the power plant less efficient which can result in a boiler blow-down whereby the unit comes off line for essential maintenance.
Should this happen then the individual power plant is not operating at peak performance and generates less power - potentially the national grid has to source this shortfall from elsewhere to supply the consumer - and the owners of the power plant lose important revenue making such a chain of events commercially undesirable and best prevented if possible.
Voltammetry (VA) can be used to quickly ensure that the iron content poses no risk to the continual and safe operation of a power plant using a bromate/tri-ethanolamine electrolyte that buffers the solutions to pH > 9.5.
Voltammetry is a technique first introduced by Heyrovsky in 1922, but in recent years outstanding advances from Metrohm have been made with regard to the measuring cell and PC control system so that today VA represent a clean, reliable and robust technique that gives quality results day in day out.
A typical instrument configuration would be the Metrohm MVA1 Voltammetry system.
Full control of the instrument is through the PC and a result of this key corrosion indicator (iron) obtained in little over 10 minutes.
A major advantage of VA is the use of standard addition for calibration which eliminates many of the matrix effects seen in spectroscopic techniques.
Voltammetry has officially established methods and two final points of importance are the running costs themselves which are minimal with no expensive gases required and the footprint which occupies less than a metre, essential in laboratories or control rooms where space is at a premium.
To learn more about the analysis of iron in power plant waters using Voltammetry please contact the Metrohm VA Support Team will be delighted to provide further application information. Request a free brochure from Metrohm UK ...