Metrohm explains Karl Fischer Oven Method
Metrohm's Karl Fischer (KF) Oven Method determines the water content of pharmaceuticals that are not amenable to direct KF titration because they undergo detrimental side reactions, for example.
Water determinations carried out with several pharmaceuticals show that the values obtained with the KF Oven Method all lie within the ranges specified in the European Pharmacopoeia.
Water determination in pharmaceuticals using an automated Karl Fischer Oven Technique
The automated KF Oven Technique enhances the sample throughput and repeatability of the results.
The quality and shelf life of active pharmaceutical ingredients depend primarily on the content of adsorbed (surface water) or bound water (water of crystallisation).
Although the fourth edition of the European Pharmacopeia (2002) describes direct KF titration as the most important method for determining water content, in some cases, where substances undergo side reactions with the KF reagents or where they release their water only at high temperatures, the European Pharmacopeia recommends quantification via the loss on drying (LOD) in a drying cabinet or desiccator.
This method also includes all volatile compounds (such as impurities or degradation products) released at higher temperatures.
Depending on the water content of the sample, the KF titration can be performed either volumetrically for higher water contents or coulometrically for lower water contents.
Considering the relatively small amount of water in most pharmaceuticals, samples are analysed coulometrically.
The calculation of the sample's water content also involves the determination of the humidity in the sample vessel and the moisture adhering to the vessel surface, vial cap and septum.
Three empty but sealed sample vials are analysed under the same conditions as the particular sample.
The resulting mean value is stored as a blank value and automatically taken into account in the calculation.
Next, a threefold determination for each compound is performed.
The accuracy of the KF titrations is checked at regular intervals with a certified KF oven standard.
The optimum oven temperature for driving off the water should ensure complete water extraction in a reasonable time, excluding the decomposition of the sample.
This means that the oven temperature should be chosen as high as possible to ensure short determination times, but should still be 20C to 30C below the decomposition temperature.
The oven temperatures for unknown samples are determined on the basis of the water-release curves that were recorded for all investigated pharmaceuticals in the 50-250C temperature range.
In contrast to the range of the recovery rates for the LOD specified by the European Pharmacopoeia (60-120 per cent), the KF Oven Method provides recovery rates between 90 per cent and 110 per cent.
The water content is determined with excellent repeatability in less than 12 minutes, according to the company.
Not what you're looking for? Search the site.
Tel +44 1928 579 600
-
Automatic water determination in pharmaceuticals
The Karl Fischer oven method of water determination, using the Metrohm 774 Oven Sample Processor, measures water content and avoids many of the normal problems encountered -
Metrohm unveils free water-analysis brochure
Metrohm has introduced a free brochure, entitled 'Water Analysis', which is designed to help users choose the best methods to determine water substances. -
Metrohm discusses electrode applications
Metrohm provides a review of electrodes and their applications. -
Analyser determines TOC level in process water
Metrohm has revealed that the Quick TOC loop can be used for the online determination of TOC in process water in the pharmaceutical industry. -
Aquatrode Plus controls quality of bioethanol
The Metrohm Aquatrode Plus is a combined pH glass electrode with ground-joint diaphragm, recommended for measuring pH in organic solvents.
Browse by category
- Plant Instrumentation (6589)
- Valves and manifolds (596)
- Valve positioners and actuators (210)
- Level and tank contents instrumentation (511)
- Temperature measurement: Conventional (314)
- Temperature measurement: IR techniques (294)
- Pressure measurement systems (443)
- Weighing systems, Strain gauges (555)
- Liquid Analysis: Titration, HPLC, IC (340)
- Gas monitoring and analysis (351)
- Environmental sensors and sampling systems (59)
- Optical sensors and vision systems (247)
- Fire, smoke and gas leak detection (140)
- Condition monitoring and vibration analysis (417)
- Position, speed and torque measurement (487)
- Electrostatics, lightning and earthing (96)
- Medical, biological and bacteriological sensors (8)
- Humidity and moisture sensing systems (219)
- Powder, droplet and particle characterisation (329)
- Corrosion Monitoring (13)
- Liquid Analysis: Spectroscopy, Colour (291)
- Noise and acoustics measurement (70)
- Liquid Analysis: Turbidity, Particles (29)
- Liquid Analysis: DO, Ozone, BOD, TOC (64)
- Liquid Analysis: pH, Conductivity, ISE (206)
- Liquid Analysis: Viscosity, Density (146)
- Positioners, servo drives (151)
- Process Control and Management (3038)
- Communications and HMI (2977)
- Process Plant and Equipment (7474)
- Processing Industry Events and Services (3625)
- News from specific industry sectors (2846)
- Environment, Pollution + Green Power (902)
- Flow Measurement (947)
- Water Industry News (1552)