Three CEM systems have been installed on the Daldowie sludge drying plant near Glasgow, monitoring the gaseous effluent from three Regenerative Thermal Oxidiser emission stacks
The Daldowie Fuel Plant at Daldowie, near Glasgow, is operated by SMW, a wholly owned subsidiary of Scottish Power Generation It includes one of Europe's largest sludge drying plants, which processes the sewage sludge from the Greater Glasgow population of approximately one million
This article was originally published on Processingtalk on 3 Jul 2008 at 8.00am (UK)
Related stories
Measurement of nitrogen in argon and helium
Exceptional precision, proven reliability and impressive connectivity makes the Kontrol Analytik k2001 on-line analyser the best choice for the measurement of trace nitrogen in argon or helium
Oxygen analyser in catalyst production SIS loop
Sud-Chemie AG is one of the world's largest manufacturers of catalysts for the chemical and petro-chemical industries: Servomex process oxygen analysers play a crucial role in the plant safety
The plant represents the very latest in sophisticated control, thermal drying and effluent treatment practice, and produces up to 65,000 tonnes a year of low odour granule.
Six dryers whose emissions feed three Regenerative Thermal Oxidisers (RTOs) each have their own emission stack.
In order to comply with strict emissions regulations set by the Environment Agency, and also to ensure that the RTOs are operating correctly, it became necessary to install a CEM (Continuous Emission Monitoring) system on each exhaust gas stack.
Further reading
Petrochemical gas analysis for CEM and safety
A major Middle Eastern ethylene production (EO/EG) facility is using Servomex gas analysers for CEM (continuous emissions monitoring) and safety applications
Six combustion gas analysers help BP Chemicals
Servomex oxygen and combustibles analysers supplied for a refurbishment project at the BP Chemicals acetic acid production plant in Hull have contributed to a significantly reduced fuel consumption
MCERTS approval for N2O measurements
MCERTS approval makes Servomex infra-red gas analysers attractive to companies seeking to reduce Nitrous Oxide emissions and claim credits under the Kyoto protocol
Three such systems were installed, incorporating Servomex 4900 series gas analysers, Servomex 2700 series gas analysers, Servomex 4995 series sample conditioning systems and ancillary equipment.
Emissions regulations in Scotland are stricter than in England - and, indeed, many other countries - and oblige plants such as Daldowie to deliver monthly reports to demonstrate that emissions limits have not been exceeded.
Due to the unusual nature of the Daldowie plant, it was required to meet newly introduced and strictly enforced regulations for the emission of gases and odours, as laid down by the Scottish Environment Protection Agency, SEPA.
The main area of concern was exhaust gases from the three Regenerative Thermal Oxidisers.
These pieces of equipment heat the gases released during the drying process so that the volatile organic compounds (VOCs) are broken down and burnt.
This was a particularly challenging application in terms of gas analysis, because of the moist nature of the gases, the continually varying mix of gases and emissions, the elevated temperature (around 120 degrees C), and the fact that several gases had to be monitored.
The new emissions regulations imposed on the plant required the measurement not only of VOCs but also CO and O2.
Furthermore, accuracy was essential because if the emissions limits are exceeded, the plant must be shut down; any false alarms due to inaccurate readings could therefore be extremely costly in terms of downtime and lost production.
It was already known that similar sludge treatment plants had suffered through unreliable CEM systems that had failed after being used to monitor production runs for as little as 10 hours.
SMW wanted to avoid this problem with the Daldowie plant, so the company wrote a tight specification.
Compliance with MCERTS (the monitoring certification scheme) requirements was also necessary, yet the equipment had to be reasonably priced.
Servomex was able to meet the specification with a bespoke walk-in cabinet housing three CEM systems, one per stack.
Each system comprises a Servomex 4900 series gas analyser, Servomex 2700 series gas analyser, TOC (Total Organic Carbon) analyser and a Servomex 4995 series sample conditioning system.
Because of the moisture content in the flue gases, the sensor probes and sample lines are heated to prevent the formation of condensation.
The Servomex 4900 gas analyser is designed specifically for CEM applications and is capable of measuring O2, CO, CO2, NO and SO2, though in the Daldowie plant only CO and O2 are being monitored.
Benefits provided by the 4900 analyser include a low cost of ownership, low maintenance requirement and easy integration with other systems.
While oxygen is measured using a paramagnetic sensor cell, infrared technology is used for carbon monoxide.
A second O2 reading is obtained using a Servomex 2700 series analyser.
The benefits provided by the 2700 include a continuous measurement of O2 and/or combustibles, high temperature operation up to 1750C, low operating/maintenance costs and is fully field maintainable.
The reason for taking two O2 readings is that the Servomex 4900 measures O2 on a dry basis and the Servomex 2700 measures O2 on a wet basis using a zirconium oxide sensor.
The 02 measurements are used in order to derive the moisture content of the gas which is used to correct the TOC measurement from a wet to dry basis, and also to convert the measured gas concentrations to reference conditions of 11% O2 as required by SEPA.
Bryan England, a consultant to SMW, comments: "Considering the complexity of the analysis being undertaken, the Servomex equipment has performed very well: as the sampled gas is both hot and wet, it is important to have a system that can handle it in such a form.
The system provided by Servomex has proved to be reliable and we have a high degree of confidence in the results it produces.
Without the Servomex CEM system SMW would find it much more difficult to comply with the stringent emissions regulations here in Scotland.
We enjoyed excellent service from Servomex, right from the initial enquiry through installation and commissioning.